A complete workflow for fluorescent labelling of extracellular vesicles (EVs) for downstream analytics and functional studies.

Fluorescent membrane dyes provide a universal strategy for detecting EVs across a wide range of analytical platforms. However, successful EV staining requires careful optimisation of dye concentration, purification and appropriate controls to minimise background fluorescence and generate reliable, reproducible data.

Applications of EV staining

Fluorescent labelling provides a versatile approach for detecting and tracking EVs across analytical and functional studies. Because EVs are nanoscale and lack a universally expressed marker, fluorescence can provide an additional means of identifying labelled EV populations within complex nanoparticle samples [1].

For analytical applications, fluorescent labelling can support EV detection, quantification and multiparametric analysis using techniques such as flow cytometry and fluorescence NTA. In functional studies, labelled EVs can be tracked during cellular uptake, biodistribution and in vivo studies, helping researchers investigate EV localisation and interactions with recipient cells.

Application Purpose
Flow cytometry Quantification and phenotyping
Fluorescence NTA Fluorescence particle detection
Confocal microscopy Cellular uptake
Super-resolution microscopy Single EV imaging
Biodistribution In vivo tracking
Functional assays Cell interactions

Micelle and aggregate formation

Many lipophilic membrane dyes self-associate to form fluorescent micelles or aggregates that overlap with EVs in size. These particles may be detected by single-particle analytical techniques, including flow cytometry and nanoparticle tracking analysis, and can be taken up by cells, leading to false-positive EV signals [1,2].

Background fluorescence from excess unbound dye

Even when dye aggregates are absent, excess unbound fluorescent dye remaining in the sample can generate substantial background fluorescence [1]. Because EV fluorescence is relatively weak, this background signal can mask labelled EVs, reducing assay sensitivity and making data more difficult to interpret.

Dye transfer between membranes

Some lipophilic dyes can diffuse between lipid membranes after staining (Figure 1). During uptake or biodistribution experiments, fluorescence may therefore arise from dye transfer to cell membranes or intracellular organelles rather than true EV uptake [3].

Minimising dye transfer is important for generating reliable conclusions from functional studies.

Figure 1: Comparison of membrane dye transfer between conventional carbocyanine dyes and Vesi-Dye LMB. A) 130 nm liposomes labelled with DiD or DiO were mixed at varying DiD:DiO ratios (0:5, 2:3, 3:2 and 5:0) and analysed by CytoFLEX flow cytometry following 1 h incubation. The presence of dual-positive liposomes indicates transfer of carbocyanine dyes between lipid membranes. B) In contrast, liposomes labelled with Vesi-Dye LMB were analysed after both 1 h and 24 h incubation and showed no detectable dye transfer, demonstrating stable membrane labelling and preserving the identity of the two liposome populations over time.

Photobleaching and signal stability

Fluorescence intensity can decrease over time due to photobleaching or dye quenching. This is particularly important for long-term live-cell imaging, uptake studies and in vivo biodistribution experiments, where stable fluorescence is essential for accurate interpretation.

Figure 2: Uptake of Vesi-Dye LMB-labelled extracellular vesicles (EVs). HEK-293T cells were treated with 20 ug of PC-3-derived EVs stained with Vesi-Dye LMB-600 (red) after 24 hours of incubation. Nuclei stained with SYTO Far red (blue) (THUNDER Imager, Leica).

Instrument compatibility

Not all fluorophores are compatible with every analytical platform. Before beginning an EV staining experiment, ensure the selected dye matches the available laser wavelengths and emission filters on your flow cytometer, microscope or nanoparticle analysis instrument. Choosing a spectrally compatible fluorophore maximises sensitivity and helps avoid unnecessary optimisation.

 

Figure 3. Fluorescent excitation and emission spectra of the Vesi-Dye LMB range (Vesiculab).
Explore Vesi-Dye LMB

How can excess dye or antibody be removed?

Following fluorescent labelling, excess dye or fluorescently conjugated antibodies should be removed before downstream analysis.

Free dye or fluorescent conjugated antibodies can:

  • Increase background fluorescence
  • Reduce assay sensitivity
  • Produce false-positive signal

Effective removal of excess dye or antibody improves the signal-to-background ratio, allowing true EV fluorescence to be distinguished with greater sensitivity and confidence.

Common methods for removing excess dye or antibody

Common purification methods such as ultracentrifugation, conventional size exclusion chromatography (SEC) and ultrafiltration can be effective but may also result in EV loss, sample dilution or lengthy processing times.

A rapid, gentle purification method that maintains EV recovery is therefore an important component of any fluorescence-based EV workflow.

Why choose Vesi-SEC?

Vesi-SEC spin-columns provide rapid purification of labelled EVs by removing excess fluorescent dye or antibody in a single one-minute low-speed centrifugation step. Unlike conventional SEC workflows, Vesi-SEC maintains high EV recovery without sample dilution, helping generate cleaner fluorescence signals while preserving valuable samples for downstream analysis.

A rapid, gentle purification method that maintains EV recovery is therefore an important component of any fluorescence-based EV workflow.

  • Removes excess fluorescent dye and antibody
  • One-minute purification
  • Non-dilutive
  • >90% particle recovery
  • Compatible with a wide range of downstream fluorescence applications
Figure 4: Jurkat-derived EVs labelled with Bodipy FL Maleimide, before and after clean-up by Vesi-SEC Micro (NanoAnalyzer, NanoFCM). Removal of free dye with Vesi-SEC Micro decreased the baseline (bottom) and increased the detected positive population from 54.3% to 66.4% (top).

A recommended workflow for fluorescence-based EV analysis

Successful fluorescence-based EV analysis depends on more than selecting an appropriate membrane dye. Optimising staining conditions, removing excess dye and validating the final workflow are all essential for generating reliable, reproducible data.

Why choose Vesi-Dye LMB?

  • No aggregate formation
  • Stable membrane labelling
  • Generates bright, stable signal
  • Available in variety of colours
  • Compatible with rapid, non-dilutive Vesi-SEC purification
Figure 4: Vesi-Dye LMB-490 efficiently labels PC3-derived EVs and liposomes (right), with minimal dye aggregation in solution (left).
Figure 5: PC-3-derived EVs were stained with Vesi-Dye LMB-490 and analysed on the AmnisⓇ ImageStreamⓇX Mk II Imaging Flow Cytometer (labelled PC-3-derived EVs were excited using a 405 nm violet laser and emission was collected on Channel 2 480-560 nm filter).
Learn more about Vesi-Dye LMB

How can EV staining workflows be validated?

Use fluorescent EV reference materials to:

  • Optimise detector settings
  • Validate staining
  • Compare experiments
  • Troubleshoot workflows

Reference materials provide confidence that fluorescence measurements reflect biological differences rather than variation introduced during staining or analysis.

Explore Vesi-Ref

Practical tips for successful EV staining

  • Filter buffers (0.22 μm filter)
  • Optimise dye concentration by titration for every new dye and EV sample type
  • Consider the particle:dye ratio
  • Remove excess dye prior to analysis
  • Optimise incubation conditions
  • Include appropriate controls

Recommended controls

Appropriate controls are essential for distinguishing EV-associated fluorescence from background signal and labelling artefacts. Current MISEV guidelines recommend the use of label-only and procedural controls when fluorescently labelling EVs, particularly where post-labelling purification is required [1, 4].

Control Purpose
Buffer only Background
Dye only Free dye
EV only Autofluorescence
Purified EV + dye Final sample

Practical tips for successful EV staining

FAQs

Excess fluorescent dye, or fluorescently-conjugated antibody, remaining after EV staining can increase background fluorescence and make it more difficult to distinguish labelled EVs. This is particularly important for EV analysis because individual vesicles generate relatively weak fluorescence signals compared with cells.

Removing excess dye after staining improves the signal-to-background ratio and helps ensure that fluorescence detected during downstream analysis is associated with the labelled nanoparticle preparation rather than residual unbound dye. Vesi-SEC provides a rapid, non-dilutive workflow for removing excess fluorescent dye prior to analysis.

EVs isolated from plasma can be fluorescently labelled, but plasma presents additional challenges because it contains abundant lipoproteins and other lipid-containing particles that may co-isolate with EVs and can also interact with membrane dyes.

For fluorescence-based analysis of plasma EVs, effective upstream purification and appropriate controls are therefore particularly important. Vesiculab recommends using the Vesi-Omics, an all-in-one plasma EV isolation kit, prior to fluorescent labelling, followed by optimisation of the staining conditions and removal of excess dye using Vesi-SEC.

Yes. Vesi-Dye LMB membrane dyes can be used to fluorescently label liposomes as well as extracellular vesicles and other lipid membrane-based nanoparticles. This makes them suitable for fluorescence-based characterisation, uptake studies and other applications where tracking lipid nanoparticles is required.

Following labelling, excess unbound dye can be removed using Vesi-SEC before downstream analysis.

Vesi-Dye LMB is available in a range of fluorescence colours to support different analytical platforms and optical configurations. Labelled nanoparticles can be analysed using compatible fluorescence-based techniques including flow cytometry, fluorescence nanoparticle tracking analysis (F-NTA) and fluorescence microscopy.

The appropriate Vesi-Dye LMB should be selected according to the excitation lasers and emission filters available on the instrument.

View the Vesi-Dye LMB excitation and emission spectra

EV staining should be validated using appropriate experimental controls to distinguish nanoparticle-associated fluorescence from autofluorescence, free dye, dye-derived particles and instrument background.

Useful controls include unstained EVs, dye-only controls and dye-only controls subjected to the same purification procedure as the stained EV sample. Nanoparticle reference materials can provide additional controls for instrument setup, fluorescence detection, workflow validation and ongoing quality control.

The Vesi-Ref portfolio includes biological EV, liposome and multi-fluorescent reference materials designed to support the standardisation and validation of EV and nanoparticle workflows.

References

  • 1. Welsh, J.A., Goberdhan, D.C.I., O’Driscoll, L., et al. (2024). Minimal information for studies of extracellular vesicles (MISEV2023): from basic to advanced approaches. Journal of Extracellular Vesicles, 13, e12404. DOI: 10.1002/jev2.12404.
  • 2. Pužar Dominkuš, P., Stenovec, M., Sitar, S., Lasič, E., Zorec, R., Plemenitaš, A., Žagar, E., Kreft, M. and Lenassi, M. (2018). PKH26 labeling of extracellular vesicles: Characterization and cellular internalization of contaminating PKH26 nanoparticles. Biochimica et Biophysica Acta (BBA) – Biomembranes, 1860(6), 1350–1361. DOI: 10.1016/j.bbamem.2018.03.013.
  • 3. Simonsen, J.B. (2019). Pitfalls associated with lipophilic fluorophore staining of extracellular vesicles for uptake studies. Journal of Extracellular Vesicles, 8(1), 1582237. DOI: 10.1080/20013078.2019.1582237.
  • 4. Welsh, J.A., van der Pol, E., Arkesteijn, G.J.A., et al. (2020). MIFlowCyt-EV: a framework for standardized reporting of extracellular vesicle flow cytometry experiments. Journal of Extracellular Vesicles, 9(1), 1713526. DOI: 10.1080/20013078.2020.1713526