Introduction

Extracellular vesicle (EV) flow cytometry enables the rapid, multiparametric analysis of individual EVs, providing information on particle concentration, fluorescence and surface-marker expression. However, EV analysis presents challenges that are not typically encountered when analysing cells.

EVs are small, heterogeneous and have a relatively low refractive index. Consequently, their light-scatter signals can fall close to the detection limits of conventional flow cytometers. Fluorescence signals from conjugated antibodies may also be weak because individual EVs carry fewer epitopes than cells. Instrument configuration, detector sensitivity, triggering thresholds and gating therefore have a substantial influence on which events are detected and ultimately classified as EVs [1-3].

Biological EV samples can be difficult to use for initial instrument optimisation because their size, concentration and fluorescence intensity are often unknown. Reference materials provide a defined and reproducible population against which instrument performance can be assessed. They can be used to establish nanoparticle detection, optimise acquisition settings and support consistent EV gating before experimental samples are analysed.

Reference Materials for Extracellular Vesicle Flow Cytometry

Different reference materials support different aspects of the EV flow cytometry workflow.

Vesi-Ref Rainbow

Vesi-Ref Rainbow is a monodisperse, fluorescent nanoparticle reference material (~140 nm) that can be detected across multiple fluorescence channels. It provides a consistent nanoparticle population for confirming that nanoscale particles are detectable above instrument background, optimising scatter and fluorescence detector settings, assessing detection across different fluorescence channels, monitoring instrument performance between experiments and supporting the transfer of workflows between flow cytometry platforms.

Vesi Ref CD63 mNG

Vesi-Ref CD63-mNG is an extracellular vesicle isolate derived from HEK 293 cells engineered to express the fluorescent protein mNeonGreen in the presence of the CD63 tetraspanin protein. Unlike synthetic particle standards, it provides a fluorescent reference population with EV-relevant physical and optical properties.

Vesi Ref CD63 mNG can be used to confirm the detection of fluorescent EVs, establish a defined EV gate and monitor consistency between experimental runs. The use of EV-based reference materials is particularly valuable because the light-scattering behaviour of nanoparticles depends on both particle size and refractive index. Synthetic particles do not necessarily generate the same scatter signal as similarly sized biological EVs [3,4].

Optimising Nanoparticle Detection Across Flow Cytometry Platforms

Vesi‑Ref Rainbow provides a defined, multicolour nanoparticle population for confirming nanoscale particle detection and assessing fluorescence performance across different optical configurations. To demonstrate its applicability across flow cytometry platforms, Vesi‑Ref Rainbow was analysed using a CytoFLEX S, an ImageStream imaging flow cytometer and a NanoAnalyzer.

Detection Using the CytoFLEX S

A water control was first acquired to establish the background signal generated by the diluent and instrument. Vesi‑Ref Rainbow was then analysed using the same acquisition settings. A distinct nanoparticle population was resolved from the background based on side scatter and fluorescence, enabling a Vesi‑Ref Rainbow gate to be established (Figure 1).

Within the gated population, Vesi‑Ref Rainbow was detected across the PerCP, APC, APC‑A700 and APC‑A750 channels. More than 92% of gated events were positive in each of the four fluorescence channels shown, demonstrating broad fluorescence detection across the available optical configuration.

 

Figure 1. Multicolour characterisation of Vesi‑Ref Rainbow by flow cytometry using a CytoFLEX S (Beckman Coulter). The upper panels show the H₂O control and Vesi‑Ref Rainbow particles plotted by VSSC against FITC fluorescence, with the Vesi‑Ref Rainbow population gated for subsequent analysis. Histograms show fluorescence of the gated population across the PerCP, APC, APC‑A700 and APC‑A750 channels. Credit: Professor Andrew Devitt, Aston University.

Detection Using ImageStream Imaging Flow Cytometry

Vesi‑Ref Rainbow was also analysed using an ImageStream (Cytek) imaging flow cytometer (Figure 2). Defined, monodisperse fluorescence intensity distributions were observed in the green, yellow and red fluorescence channels. Representative images confirmed multichannel detection of individual Vesi‑Ref Rainbow particles, alongside corresponding side-scatter signals. Imaging flow cytometry provides the additional benefit of visualising the fluorescent signal from individual particles (Figure 2, bottom panel) [6].

Figure 2. Characterisation of Vesi‑Ref Rainbow using imaging flow cytometry (ImageStream, Cytek). Histograms show green fluorescence intensity in Ch02 (480–560 nm), yellow fluorescence intensity in Ch03 (560–595 nm) and red fluorescence intensity in Ch05 (642–745 nm). Representative images show fluorescence in Ch02, Ch03 and Ch05 alongside side scatter in Ch06. Credit: Megan Bannister and Dr Joshua Price, University of Birmingham.

Detection Using the NanoAnalyzer

Vesi Ref Rainbow was also analysed using a NanoAnalyzer (NanoFCM), a flow cytometer designed for single-particle analysis at the nanoscale. Side scatter was used to trigger particle detection, allowing the reference population to be resolved clearly from the instrument and diluent background.

Following excitation with the 488 nm and 640 nm lasers, Vesi Ref Rainbow produced distinct fluorescence signals in the green and red detection channels, respectively (Figure 3). Approximately 97% of gated particles were classified as fluorescence-positive in each channel under the acquisition and analysis conditions used. This defined, dual-colour population provides a practical reference for confirming small-particle detection, assessing fluorescence-channel performance and establishing suitable acquisition settings before experimental EV samples are analysed.

Figure 3: Characterisation of Vesi-Ref Rainbow using the NanoAnalyzer (NanoFCM). Vesi-Ref Rainbow particles were excited using 488 nm and 640 nm lasers, and the corresponding side scatter (SS), FITC (525/40), and APC (670/30) signals were detected. Side scatter (SS) was used as the trigger channel. Vesi-Ref Rainbow was diluted 1:2000 in ddH₂O to achieve an optimal event rate. Credit: Professor Andrew Devitt, Aston University.

Establishing an EV Gate Across Flow Cytometry Platforms

While Vesi‑Ref Rainbow supports general nanoparticle detection and fluorescence optimisation, Vesi‑Ref CD63‑mNG provides a biologically relevant fluorescent EV population that can be used to establish an EV gate.

EV Detection Using ImageStream

Vesi‑Ref CD63‑mNG was analysed using imaging flow cytometry. Individual events were visualised in brightfield and in the green fluorescence channel, confirming the detection of fluorescent EVs. A distinct population was identified using side scatter and Ch02 green fluorescence intensity (Figure 4). A fluorescence gate applied to the total population classified 89.83% of detected events as mNeonGreen-positive under the acquisition and analysis conditions used.

Figure 4: Establishment of a fluorescent EV gate using Vesi Ref CD63 mNG on the ImageStream platform (Cytek). Representative brightfield (Ch01) and green fluorescence (Ch02) images of Vesi-Ref CD63-mNG particles (left). SSC versus green fluorescence intensity (480–560 nm) density plot showing the fluorescent particle population (centre). Histogram of green fluorescence intensity demonstrating that 89.83% of detected particles are mNeonGreen-positive (mNG⁺) (right). Credit: Megan Bannister and Dr Joshua Price, University of Birmingham.

EV detection Using the CytoFLEX Nano

Vesi‑Ref CD63‑mNG was also evaluated on the CytoFLEX Nano in collaboration with Beckman Coulter Life Sciences. Following use of filtered PBS to characterise the background, the reference material produced a defined particle population based on violet side scatter and mNeonGreen fluorescence. Differences in scatter and fluorescence intensity could also be used to distinguish subpopulations within the Vesi‑Ref CD63‑mNG sample, demonstrating its application for establishing and evaluating EV detection regions on the CytoFLEX Nano.

Figure 5: Establishment of a fluorescent EV gate using Vesi Ref CD63 mNG on the CytoFLEX Nano platform (Beckman Coulter). A) PBS buffer (20 nm filtered) used to set VSSC1-H gating threshold. B) Population distribution of Vesi-Ref CD63-mNG EVs in VSSC1-H channel. C) Vesi-Ref CD63-mNG gated into small and large populations by VSSC1-H scatter and mNG intensity.

Further details on instrument setup, acquisition and analysis are available in the full Vesi‑Ref CD63‑mNG CytoFLEX Nano application note, developed in collaboration with Beckman Coulter Life Sciences. Read the full application note here

EV Detection Using the NanoAnalyzer

Vesi‑Ref CD63‑mNG was also analysed using the NanoAnalyzer. A defined mNeonGreen-positive EV population was resolved above the background, enabling an EV gate to be established in the green fluorescence channel (Figure 6).

The detection of Vesi‑Ref CD63‑mNG using both imaging and dedicated small-particle flow cytometry demonstrates how an EV-based reference material can support EV gating across platforms with different detection technologies.

Figure 6. Establishment of a fluorescent EV gate using Vesi Ref CD63 mNG on the NanoAnalyzer platform (NanoFCM). A) Vesi Ref Rainbow was analysed using side-scatter triggering (488nm). Fluorescence was detected in green (525/40) and red (670/30) channels. Vesi-Ref Rainbow was diluted 1:2000 in ddH₂O to achieve an optimal event rate. B) Vesi-Ref CD63-mNG allowed a defined EV population to be resolved above the background, allowing an EV gate in the green channel to be established. Vesi-Ref CD63-mNG was diluted 1:100 in PBS to achieve an optimal event rate. Credit: Professor Andrew Devitt, Aston University.

Because Vesi‑Ref CD63‑mNG particles are EVs, the detected population provides an EV-based reference rather than a synthetic particle gate. However, experimental EV populations may differ in size, fluorescence intensity, marker abundance and scatter. The reference gate should therefore provide a starting point for analysis, supported by buffer-only, unstained EV, reagent-only and fluorescence controls [1,2].

A Practical EV Flow Cytometry Workflow

Reference materials provide a stable starting point for instrument setup, but they form one part of a complete extracellular vesicle flow cytometry workflow.

1. Establish Instrument Background
Run clean, particle-free buffer using the intended acquisition settings. This identifies instrument noise and any particles introduced by the diluent or fluidics.

2. Confirm Nanoparticle Detection
Run Vesi-Ref Rainbow to confirm that a defined nanoscale population can be distinguished from the background. Optimise the relevant scatter and fluorescence detectors as required.

3. Establish an EV Gate
Run Vesi Ref CD63 mNG to identify a defined fluorescent EV population. Adjust the green fluorescence detector and establish the EV detection region using the reference population.

4. Fix the Acquisition Settings
Once the reference populations are reproducibly detected, retain the same acquisition settings for the associated controls and experimental samples. This reduces technical variation introduced by adjusting detector settings separately for each sample.

5. Analyse the Experimental Controls
Run the relevant buffer, unstained EV, reagent-only and fluorescence controls. Serial dilution should also be considered to identify coincident or “swarm” detection, in which multiple EVs are recorded as a single event [1,2,7].

6. Acquire the Experimental EV Samples
Apply the established acquisition framework to the experimental samples. The Vesi Ref CD63 mNG gate can provide a starting EV gate, while the accompanying controls are used to identify background and validate the analysis regions for the biological sample.

Supporting More Reproducible Extracellular Vesicle Flow Cytometry

Reference materials provide defined, reproducible populations against which flow cytometer performance can be assessed, reducing reliance on unknown and heterogeneous biological samples during initial instrument setup.

Vesi‑Ref Rainbow supports confirmation of nanoparticle detection, optimisation of multiple fluorescence channels and monitoring of instrument performance over time. Vesi‑Ref CD63‑mNG complements this by providing a fluorescent EV population for establishing an EV gating strategy and optimising settings relevant to green fluorescent EV analysis. Together, the materials provide a practical framework for standardising acquisition settings and improving consistency between experiments and across flow cytometry platforms.

Reference materials do not replace appropriate experimental controls or account for the variability of biological EV samples. Rather, they complement these controls by supporting consistent instrument setup and more reproducible analysis of experimental EV samples.

FAQs

Reference materials provide a consistent population for confirming nanoparticle detection, optimising scatter and fluorescence settings, establishing gates and monitoring instrument performance before experimental EV samples are analysed.

Yes, Vesi‑Ref CD63‑mNG consists of fluorescent CD63‑mNeonGreen EVs, providing a defined EV population that can be used to establish an EV gate. Appropriate experimental controls should still be used to validate the gate for each biological sample.

Vesi-Ref Rainbow is a multicolour fluorescent nanoparticle reference material for assessing nanoparticle detection and fluorescence performance. Vesi Ref CD63 mNG is an EV-based fluorescent reference material designed to support fluorescent EV detection and EV gating.

Vesi Ref materials can support method development across compatible conventional, imaging and nano-flow cytometry platforms. Detection will depend on the instrument’s lasers, optical filters, sensitivity and acquisition settings.

Biological EV samples are heterogeneous and their expected scatter and fluorescence intensities may be unknown. A reference material provides a consistent population that can be used to optimise the instrument before experimental samples are introduced.

No. Reference materials help establish instrument performance and acquisition settings, but buffer, unstained EV, reagent-only, fluorescence and dilution controls remain important for identifying background, non-EV events and coincident detection.

References

  1. Welsh JA, Van Der Pol E, Arkesteijn GJA, et al. MIFlowCyt-EV: a framework for standardized reporting of extracellular vesicle flow cytometry experiments. Journal of Extracellular Vesicles. 2020;9:1713526. https://doi.org/10.1080/20013078.2020.1713526
  2. Welsh JA, Goberdhan DCI, O’Driscoll L, et al. Minimal information for studies of extracellular vesicles (MISEV2023): From basic to advanced approaches. Journal of Extracellular Vesicles. 2024;13:e12404. https://doi.org/10.1002/jev2.12404
  3. Lannigan J, Erdbruegger U. Extracellular vesicle flow cytometry analysis and standardization. Frontiers in Cell and Developmental Biology. 2017;5:78. https://doi.org/10.3389/fcell.2017.00078
  4. van der Pol E, Sturk A, van Leeuwen TG, Nieuwland R, Coumans FAW. Standardization of extracellular vesicle measurements by flow cytometry through vesicle diameter approximation. Journal of Thrombosis and Haemostasis. 2018;16:1236–1245. https://doi.org/10.1111/jth.14009
  5. Welsh JA, Van Der Pol E, Bettin BA, et al. Towards defining reference materials for measuring extracellular vesicle refractive index, epitope abundance, size and concentration. Journal of Extracellular Vesicles. 2020;9:1816641. https://doi.org/10.1080/20013078.2020.1816641
  6. Görgens A, Bremer M, Ferrer-Tur R, et al. Optimisation of imaging flow cytometry for the analysis of single extracellular vesicles by using fluorescence-tagged vesicles as biological reference material. Journal of Extracellular Vesicles. 2019;8:1587567. https://doi.org/10.1080/20013078.2019.1587567
  7. van der Pol E, van Gemert MJC, Sturk A, Nieuwland R, van Leeuwen TG. Single versus swarm detection of microparticles and exosomes by flow cytometry. Journal of Thrombosis and Haemostasis. 2012;10:919–930. https://doi.org/10.1111/j.1538-7836.2012.04683.x